Chapter 9 — Attentuated Total Reflection Spectroscopy: Physics, Approximations, and Best Practices

This chapter illustrates characteristic features commonly encountered in practical ATR spectroscopy, such as non-linear intensity behavior, peak shifts, and band distortions. These effects are not artifacts and do not indicate experimental shortcomings. Instead, they arise because ATR measurements probe the optical response of a sample through reflection at an interface, where electromagnetic boundary conditions and dispersion directly shape the observed spectra.

Importantly, this does not imply that transmission spectroscopy provides a geometry-independent or fundamentally more “correct” reference. Transmission spectra are themselves affected by optical phenomena such as reflection losses, interference effects, and dispersion-related non-linearities, which are often suppressed only under favorable experimental conditions. ATR spectroscopy differs mainly in that these optical contributions are more explicit and therefore cannot be ignored.

In the following, ATR spectra are shown in the commonly used ATR absorbance representation for practical convenience and familiarity. This representation should be understood as a conditional, non-linear transformation of reflectance, not as a direct measure of molecular absorption. The purpose of this example is therefore not to correct or eliminate such effects, but to make geometry-induced behavior visible and tangible.

ATR spectra are often treated as if they were directly comparable to transmission absorbance spectra. The following historical clarification explains why this intuition arose, when it is justified, and why it must eventually break down.

Early theoretical interpretations of ATR spectroscopy sought to explain why ATR spectra often resemble transmission absorbance under many practical conditions. Harrick (63) approached this problem using a local-field picture, in which the electromagnetic field at the surface of the internal reflection element is assumed to interact directly with molecular absorbers. Within this framework, he introduced the concept of an effective thickness, which allows ATR absorbance to be related approximately to transmission absorbance when absorption is weak.

Hansen (64) arrived at the same concept from a different direction. Starting explicitly from Fresnel reflection theory formulated in terms of the complex refractive index, he showed that, under weak-absorption conditions, ATR reflectance can be treated using a first-order approximation. In this limit, ATR absorbance becomes approximately proportional to absorption strength, multiplied by a geometry-dependent effective depth or thickness.

This effective thickness is intrinsically frequency dependent and is evaluated using the frequency-dependent real refractive indices of the sample and the internal reflection element. Dispersion therefore enters explicitly through the optical boundary conditions. Importantly, the resulting absorbance-like behavior is not fundamental. It is a controlled approximation that holds only as long as absorption remains sufficiently weak that higher-order wave-optical coupling between absorption, dispersion, and boundary conditions does not dominate the measured reflectance.

9.1 What ATR Really Measures: From “Evanescent Field” to Reflection Optics

Attenuated Total Reflection (ATR) spectroscopy is commonly introduced as a technique based on the interaction of an evanescent electromagnetic fieldwith a sample adjacent to an internal reflection element (IRE). While this description is pedagogically convenient and historically important, it obscures the physical mechanism responsible for signal formation in ATR spectroscopy and has contributed to widespread misunderstandings regarding interpretation, linearity, and correction.

A wave-optical description provides a more precise and internally consistent picture.

In a non-absorbing rarer medium, total internal reflection gives rise to an evanescent field that decays exponentially perpendicular to the interface. This field is mathematically well defined and experimentally observable. However, the moment the rarer medium becomes absorbing, the lossless picture of a purely evanescent field with no normal energy transport no longer applies. Instead, electromagnetic energy is transmitted into the sample, and the deviation from perfect reflection arises from energy transport into an absorbing medium, not from “absorption of an evanescent wave.”

Importantly, the transition between these regimes is continuous: the conventional sharp boundary between internal reflection, attenuated total reflection, and subcritical reflection becomes a continuous transition once absorption is present. The measured reflectance varies smoothly with both the angle of incidence and the optical properties of the sample. (65)
From a wave-optical perspective, ATR is therefore best understood as reflection spectroscopy under boundary conditions that suppress—but do not eliminate—transmission (see Fig. 9.1). (66)

Fresnel k-vector (wavefront-direction) map for an internal reflection element (IRE, (n_1 = 2.4)) in contact with a sample ((n_2 = 1.5)). A fan of incident plane-wave directions in the IRE meets the interface and splits into reflected and transmitted wavefront directions; line shade encodes the Fresnel amplitude. Left panels are s-polarized, right panels p-polarized; the top row is a non-absorbing sample, the bottom row an absorbing one. Below the critical angle the transmitted wave propagates into the sample; above it the wave becomes interface-bound, and absorption smooths the transition. For an absorbing sample the transmitted wavevector is complex, so the drawn directions should be read as a schematic visualization of phase propagation, not as ordinary energy-flow rays.

Fig. 9.1
Fresnel k-vector (wavefront-direction) map for an internal reflection element (IRE, \(n_1 = 2.4\)) in contact with a sample (\(n_2 = 1.5\)). A fan of incident plane-wave directions in the IRE meets the interface and splits into reflected and transmitted wavefront directions; line shade encodes the Fresnel amplitude. Left panels are s-polarized, right panels p-polarized; the top row is a non-absorbing sample, the bottom row an absorbing one. Below the critical angle the transmitted wave propagates into the sample; above it the wave becomes interface-bound, and absorption smooths the transition. For an absorbing sample the transmitted wavevector is complex, so the drawn directions should be read as a schematic visualization of phase propagation, not as ordinary energy-flow rays.

The figure illustrates, in a ray-like representation, the angular redistribution of reflected and transmitted beam components as determined by the Fresnel boundary conditions, and how absorption modifies their relative intensities. For a non-absorbing sample above the critical angle, there is no net energy transport normal to the interface into the sample. For an absorbing sample, however, energy is still coupled into the sample and dissipated there. This energy does not return to the detector as reflected radiation; the reflection is therefore no longer total, but attenuated.

Recognizing ATR as a reflection experiment has several immediate and practically important consequences:

  • The distinction between “ATR” and “internal reflection below the critical angle” is not fundamental.

  • The conventional total-internal-reflection definition of the critical angle ceases to be a sharp boundary once absorption is non-zero. In absorbing media, a more useful definition is the angle at which the real and imaginary parts of the wavevector component perpendicular to the interface become equal, \(\mathrm{Re}(k_\perp) = \mathrm{Im}(k_\perp)\), rather than the onset of total internal reflection. (65)

  • Deviations from unity reflectance must be attributed to transmission into the sample.

  • Geometry, polarization, and refractive index necessarily influence the spectrum.

A particularly instructive result is the redefinition of the critical angle in absorbing media as the point where the real and imaginary parts of the perpendicular component of the wavevector become equal. This replaces the traditional refractive-index-only criterion and highlights that ATR is governed by the full complex optical response of the sample.

To further elucidate the physical nature of ATR in absorbing samples, it is instructive—purely illustratively—to examine the spatial structure of the electromagnetic field near the IRE/sample interface as obtained from the Fresnel solution. Figure 9.2 therefore shows calculated field distributions for selected angles below and above the critical angle \(\alpha_c\), for both non-absorbing and absorbing samples. These field maps are not directly observable and should not be interpreted as measured quantities; rather, they serve to visualize the continuous evolution of the complex wavefield imposed by the boundary conditions. In particular, for absorbing samples, the near-interface field patterns below and above \(\alpha_c\) are qualitatively very similar, reflecting the fact that the normal component of the wavevector is complex at all angles. (65)

This illustrates that, in absorbing media, ATR sensitivity cannot be attributed to the sudden appearance of a distinct evanescent field at the critical angle, but instead arises from the continuous, boundary-condition-determined complex wavefield present on both sides of \(\alpha_c\). The form of these field patterns reflects the continuous coupling of light into the absorbing sample, rather than a sharp onset at the critical angle.

The Beer–Lambert limit compared with the transmittance of a finite slab, all for the same intrinsic absorption (k). The ideal curve assumes pure exponential attenuation and a baseline at unity. The coherent slab adds an interference ripple from multiple internal reflections. The incoherent slab — where thickness variations or limited coherence wash the fringes out — is perfectly smooth, yet its baseline is still depressed below unity by reflection at the two surfaces — to ((1-R)/(1+R)) in the non-absorbing limit, once incoherent multiple reflections are included (Section 7.1). Beer–Lambert therefore fails even when there is no interference at all; the depression is a pure boundary effect.

Fig. 9.2
Spatial distribution of the electromagnetic field near an internal reflection element (IRE)/sample interface calculated from the Fresnel solution for selected angles below and above the critical angle \(\alpha_c\). The upper half corresponds to the IRE (\(n_1 = 2.4\)), the lower half to the sample with complex refractive index \(\hat{n}_2 = n_2 + i k_2\); here \(n_2 = 1\), a deliberately simple illustrative value, so the critical angle differs from that in Figures 9.1 and 9.3. Results are shown for a non-absorbing sample (\(k_2 = 0\)) and for an absorbing sample (\(k_2 > 0\)). The critical angle \(\alpha_c\) is defined as the angle at which the real and imaginary parts of the wavevector component perpendicular to the interface are equal. Colors represent the real part of the electric field, \(\mathrm{Re}(E)\), in arbitrary units; a single field component is shown, which for p-polarization should not be read as the total field intensity. For absorbing samples, the near-interface field distributions below and above \(\alpha_c\) are qualitatively very similar, illustrating the continuous evolution of the Fresnel solution and the absence of a sharp transition between propagating and evanescent regimes.

The continued use of the evanescent-field description is not merely a semantic issue. It reflects the fact that, under weak-absorption conditions, ATR spectra often behave approximately like true absorbance spectra (\(A = 4\pi\,(\log_{10} e)\,\tilde{\nu}\,k(\tilde{\nu})\,d\)). In this limited regime, simplified interpretations can work remarkably well.

However, taken literally, the evanescent-field narrative encourages the misleading idea that ATR interrogates a chemically selective layer whose thickness is on the order of the wavelength. In reality, the fate of the light is decided at the interface: for a homogeneous sample the Fresnel equations fix how much radiation is reflected and how much is transmitted, governed solely by the optical contrast between the two media. There is no redistribution of the field that singles out a particular depth.

What the spectrum encodes is therefore a convolution of absorption and dispersion expressed through these boundary conditions, with relative weights that depend sensitively on the complex refractive index, oscillator strength, polarization, and angle of incidence.

Real samples complicate this picture not by introducing a simple bulk sampling depth, but by modifying the boundary-value problem — and hence the field distribution — at and near the interface. Liquids conform intimately to the IRE and approach the ideal of a single, sharp, homogeneous interface; solid samples seldom do. Imperfect optical contact, microheterogeneity, densified or chemically altered surface regions, and partial molecular orientation at the surface all modify the boundary conditions — and hence the measured spectrum — independently of any nominal penetration depth.

ATR does not measure absorption directly. It measures how the optical properties of a sample modify reflection at an interface.

This perspective does not invalidate common ATR practice. Instead, it explains:

  • why ATR spectra often resemble transmission spectra,

  • why this resemblance is conditional,

  • and why deviations become unavoidable for strong bands, high-index materials, or near-critical geometries.

Recognizing ATR as reflection spectroscopy governed by electromagnetic boundary conditions is the conceptual starting point for any physically meaningful interpretation, correction, or quantitative use of ATR data (Figure 9.3).

Normal wavevector components at the IRE–sample interface for non-absorbing and absorbing samples.

Fig. 9.3
Normal wavevector components at the IRE–sample interface for non-absorbing and absorbing samples.

Schematic illustration of the wavevector geometry at the interface between the internal reflection element (IRE) and the sample. Top row (non-absorbing sample): Below the critical angle, the normal component of the wavevector in the sample is real and energy is transmitted into the sample. Above the critical angle, the normal component becomes purely imaginary and no energy is transported normal to the interface, corresponding to total internal reflection with an evanescent field. Bottom row (absorbing sample): The normal wavevector component is complex for all angles of incidence, \(k_\perp = a + ib\), and energy is always transmitted into the sample and dissipated. The generalized critical angle marks the condition \(a=b\), separating propagation-dominated (\(a>b\)) from attenuation-dominated (\(b>a\)) behavior. Unlike the non-absorbing case, the normal wavevector component never vanishes for absorbing samples.

9.2 Electric Field Distributions, Penetration Depth, and the Limits of “Effective Thickness”

Early theoretical discussions of ATR spectroscopy focused on the question of how far the electromagnetic field extends into the sample. For non-absorbing media, this leads to the well-known expression for the penetration depth, defined as the distance at which the field amplitude decays to 1/e of its value at the interface. In this special case, the penetration depth is mathematically well defined and can be derived directly from wave optics (Figure 9.4).

ATR penetration depth versus angle of incidence for three internal reflection elements at 1000 cm⁻¹ (sample (n_2 = 1.5)). The depth diverges at each element’s critical angle (dotted) and shrinks as the angle increases. The high-index germanium element has a small critical angle and a shallow penetration depth; for ZnSe and diamond it is several times larger. The dashed line marks the common 45° working angle.

Fig. 9.4
ATR penetration depth versus angle of incidence for three internal reflection elements at 1000 cm⁻¹ (sample \(n_2 = 1.5\)). The depth diverges at each element’s critical angle (dotted) and shrinks as the angle increases. The high-index germanium element has a small critical angle and a shallow penetration depth; for ZnSe and diamond it is several times larger. The dashed line marks the common 45° working angle.

For real samples, however, absorption is unavoidable. Once absorption is present, the field no longer decays purely exponentially, and the concept of a single penetration depth loses its strict physical meaning. Because a detector provides no information about where absorption occurs, deviations from total reflection must be attributed to transmission of light into the sample. For homogeneous samples, the Fresnel equations show that the fraction of light coupled into the sample is determined by the dielectric (and, in principle, magnetic) contrast at the interface.

The desire to relate ATR spectra to familiar transmission-style absorbance — more precisely, the true absorbance defined by the absorption coefficient and path length — motivated the introduction of the effective thickness (also called effective depth). This quantity was developed independently by Harrick (63) and Hansen (64) — the wave-optical derivation following that of Milošević (67) — to answer a pragmatic question:

What thickness in transmission would produce the same absorbance as observed in ATR?

From a wave-optical standpoint, effective thickness is obtained by expanding the Fresnel reflection coefficients under restrictive assumptions, most importantly the weak-absorption approximation. In this approximation, ATR absorbance becomes approximately proportional to absorption strength, multiplied by a geometry-dependent prefactor that can be interpreted as an effective thickness.

Several important points follow immediately:

  • Effective thickness depends on polarization (s vs. p).

  • It depends on the angle of incidence.

  • In the classical weak-absorption form, it is evaluated using the real refractive indices of the sample and of the internal reflection element, with absorption entering only through the small parameter of the expansion.

  • It is frequency dependent.

  • It depends on the notion that transmission absorbance equals true absorbance (\(A = 4\pi\,(\log_{10} e)\,\tilde{\nu}\,k(\tilde{\nu})\,d\)), which is generally incorrect.

Crucially, effective thickness is not a material constant.

Because effective thickness is evaluated using the frequency-dependent real refractive indices of the sample and the internal reflection element, dispersion enters explicitly at this stage. As a result, the product of effective thickness and absorption strength often mirrors the refractive-index function of the sample (Figure 9.5).

Wavelength dependence of penetration depth and its spectral consequence. (a) Because (d_p) is proportional to wavelength, the penetration-depth/effective-thickness picture predicts a larger decay length at lower wavenumber. (b) For a sample with two equally absorbing bands, a transmission spectrum (absorbance (propto tilde{nu},k)) emphasizes the high-wavenumber band, whereas the ATR effective thickness (propto 1/tilde{nu}) compensates, leaving the bands comparable — so low-wavenumber bands are relatively enhanced in ATR.

Fig. 9.5
Wavelength dependence of penetration depth and its spectral consequence. (a) Because \(d_p\) is proportional to wavelength, the penetration-depth/effective-thickness picture predicts a larger decay length at lower wavenumber. (b) For a sample with two equally absorbing bands, a transmission spectrum (absorbance \(\propto \tilde{\nu}\,k\)) emphasizes the high-wavenumber band, whereas the ATR effective thickness \(\propto 1/\tilde{\nu}\) compensates, leaving the bands comparable — so low-wavenumber bands are relatively enhanced in ATR.

This has direct and observable consequences:

  • systematic redshifts of ATR band maxima,

  • asymmetric band shapes,

  • intensity redistribution across overlapping bands.

These effects are intrinsic consequences of the optical boundary conditions governing ATR measurements.

Effective thickness is best understood as a diagnostic and transitional concept. It captures part of the intensity redistribution responsible for ATR sensitivity and provides a useful bridge between reflection optics and transmission-style intuition.

However, two limitations are essential for best practice:

  • Effective thickness does not linearize ATR absorbance. Even when used correctly, it only approximates linearity over limited ranges of absorption strength.

  • A single, geometry-independent effective thickness does not exist. Applying one constant value across an entire spectrum introduces hidden assumptions that are rarely justified.

The widespread practice of using a single effective thickness to “correct” ATR spectra therefore risks masking physically meaningful effects rather than resolving them.

Penetration depth and effective thickness are sometimes useful approximations, not fundamental observables.

They explain why ATR spectra can resemble transmission spectra under favorable conditions, but they also explain why deviations—such as peak shifts and band distortions—are unavoidable once absorption becomes stronger or refractive-index contrast increases.

Understanding these concepts as conditional tools rather than universal corrections is essential for responsible interpretation of ATR data.

9.3 Non-Linearities in ATR Spectroscopy: Absorption, Concentration, and Dispersion

A common expectation in infrared spectroscopy is that absorbance scales linearly with molar absorption coefficient, concentration, or path length. In transmission spectroscopy this expectation can be approximately justified under well-defined conditions (e.g., sufficiently thick layers, limited reflection/interference contributions, weak bands). In ATR spectroscopy, however, such linearity is not a general property, because the measured quantity is not transmitted intensity but the frequency-dependent reflectance at an interface.

Non-linearity in ATR is not a secondary complication and not a sign of poor experimental practice. It is a direct consequence of how reflection at an interface responds to the optical properties of an absorbing medium. (68)

In an ATR experiment, the detector records the reflectance \(R(\tilde{\nu})\) at the interface between the internal reflection element (IRE) and the sample. The commonly plotted “ATR absorbance” is obtained only after a logarithmic transformation of the ratioed reflectance, \(A_{\mathrm{ATR}}(\tilde{\nu}) = -\log_{10}\!\left[R(\tilde{\nu})/R_0(\tilde{\nu})\right]\), where \(R_0(\tilde{\nu})\) is a background (reference) reflectance. Even if the sample absorption strength increased perfectly linearly with concentration, the corresponding changes in reflectance generally would not.

The underlying reason is that reflectance is governed by the Fresnel coefficients, which are non-linear functions of the sample’s complex refractive index \(\hat{n}(\tilde{\nu}) = n(\tilde{\nu}) + i k(\tilde{\nu})\), of polarization, and of the angle of incidence. As absorption increases, the complex wavevector component perpendicular to the interface changes, altering the reflection and transmission amplitudes that the boundary conditions impose at the interface, and with them the efficiency with which energy is coupled from the IRE into the absorbing sample. In other words, ATR does not measure “how much absorption occurs in a fixed path length”; it measures how the optical boundary-value problem responds to changes in \(\hat{n}\).

Three coupled contributions are particularly important:

  • Boundary-condition response near the interface: as \(k(\tilde{\nu})\) increases, the complex wavevector component perpendicular to the interface changes, altering the Fresnel solution. This changes the coupling of energy into the sample in a way that is generally non-linear in \(k\).

  • Geometry- and polarization-dependent coupling efficiency: the coupling of energy into the sample depends strongly on polarization (s vs. p) and on the angle of incidence. Especially near critical-angle geometries, small changes in optical properties can produce disproportionately large changes in reflectance. This alone can destroy Beer–Lambert-like scaling.

  • Dispersion-related feedback: absorption bands are accompanied by changes in the real refractive index \(n(\tilde{\nu})\). These dispersion changes feed back into the Fresnel coefficients and therefore into the measured reflectance. Changes in \(n(\tilde{\nu})\) modify intensity scaling and mixture additivity, even at fixed measurement geometry.

Together, these effects imply that \(A_{\mathrm{ATR}}(\tilde{\nu})\) cannot, in general, be expected to scale linearly with absorber amount, and mixture spectra cannot automatically be assumed to be linear combinations of component spectra.

Non-linearity becomes particularly obvious in systematic concentration series and in binary mixtures. Typical observable consequences include:

  • Intensity scaling is non-linear: doubling the concentration of an absorber does not necessarily double “ATR absorbance.”

  • Spectra are not strictly additive: mixture spectra cannot generally be represented as linear combinations of neat-component spectra outside a limited regime.

  • Apparent band changes can occur without chemistry: changes in spectral shape or relative band intensities may occur purely because the Fresnel boundary conditions respond non-linearly to \(\hat{n}\).

These deviations become more pronounced when (i) absorption bands are strong, (ii) the sample refractive index approaches that of the IRE, or (iii) the experiment is performed near critical-angle conditions.

A frequent misinterpretation is to attribute non-linear ATR behavior to chemical effects such as aggregation, specific interactions, or concentration-dependent transition strengths. Such chemical effects can certainly exist in real samples, but ATR optics alone is sufficient to generate pronounced non-linearities even for chemically ideal systems.

From a best-practice perspective, the distinction is essential:

  • Optical non-linearity arises inevitably from Fresnel boundary conditions and the non-linear dependence of reflectance on \(\hat{n}\).

  • Chemical non-linearity must be demonstrated independently (e.g., by complementary techniques, controlled refractive-index context, or physically grounded mixture modeling), not inferred solely from ATR intensity behavior.

Failing to separate these categories risks over-interpreting optical effects as chemistry.

Non-linearity in ATR spectroscopy is expected, not exceptional. “ATR absorbance” is a conditional representation whose apparent linearity must be demonstrated for the specific sample, geometry, and spectral range under investigation. When linearity is observed, it reflects a favorable operating regime—not a general property of ATR spectroscopy.

Understanding this point is crucial for quantitative analysis, calibration strategies, mixture interpretation, and the design of chemometric or AI models trained on ATR data (Figure 9.6).

Simulated concentration-dependent ATR spectra before and after correction. As concentration (oscillator strength) rises, the uncorrected band maxima drift to lower wavenumber (bottom) — a dispersion artifact, not a change in chemistry. A sophisticated Fresnel/KKR correction (top), here applied to a known 50:50 s/p reflectance mixture (effective polarization angle 45°) on a ZnSe element, reconstructs the complex refractive index and from it the true absorbance, restoring the band maxima to the oscillator positions (dotted lines) and recovering near-symmetric line shapes across the whole concentration range.

Fig. 9.6
Simulated concentration-dependent ATR spectra before and after correction. As concentration (oscillator strength) rises, the uncorrected band maxima drift to lower wavenumber (bottom) — a dispersion artifact, not a change in chemistry. A sophisticated Fresnel/KKR correction (top), here applied to a known 50:50 s/p reflectance mixture (effective polarization angle 45°) on a ZnSe element, reconstructs the complex refractive index and from it the true absorbance, restoring the band maxima to the oscillator positions (dotted lines) and recovering near-symmetric line shapes across the whole concentration range.

Uncorrected ATR spectra (bottom) and the corresponding corrected spectra (top) are shown for a simulated three-oscillator model whose oscillator strengths increase with concentration under identical ATR conditions. The uncorrected spectra exhibit pronounced non-linear intensity scaling, progressive red-shifts, and band-shape distortion with increasing absorber fraction, reflecting intrinsic optical effects of ATR spectroscopy and the breakdown of Beer–Lambert linearity: ATR absorbance does not scale proportionally with absorber amount, and absorbances are not strictly additive outside the linear regime.

The sophisticated Fresnel/KKR correction inverts the boundary-value problem to reconstruct the complex refractive index, and from it the true absorbance, so that the corrected band maxima return to the oscillator positions indicated by the vertical lines and the line shapes become near-symmetric across the whole series. (69),   (70) 

A caveat applies to real mixtures: a residual, composition-dependent shift can remain even after a correct ATR inversion, but it is not an ATR artifact. It arises from local-field and effective-medium behavior, which modify the dielectric response of the oscillators independently of the ATR measurement geometry and must therefore not be attributed to ATR optics. (71)

The figure illustrates that a physically grounded ATR correction recovers meaningful reference positions and line shapes, while composition-dependent shifts in mixtures belong to the material response rather than to the geometry.

9.4 Peak Shifts, Band Distortion, and Misinterpretation in ATR Spectroscopy

One of the most frequently observed, and most frequently misinterpreted, differences between ATR and transmission spectroscopy is the displacement and deformation of spectral bands. These effects are often labeled “ATR artifacts” or treated as nuisances. In reality, they are intrinsic consequences of reflection optics at an absorbing interface and carry direct information about how dispersion and boundary conditions shape the measured spectrum.

In transmission spectroscopy, apparent absorbance maxima often lie close, typically within about 1–2 cm⁻¹, to the maxima of \(\varepsilon”\) and thus to the oscillator positions. For stronger bands, the apparent maxima of \(k\) or absorbance can shift — usually blue-shifting — because the dielectric function and the complex refractive index are related non-linearly (a quadratic dependence). In contrast, in ATR spectroscopy, the observed band maximum is governed additionally by the Fresnel equations.

As a result:

  • ATR band maxima are often red-shifted relative to oscillator positions.

  • The magnitude of the shift increases with oscillator strength.

  • The effect becomes more pronounced for low refractive-index contrast between sample and IRE and for near-critical geometries.

These shifts are optical in origin and should not automatically be interpreted as changes in vibrational eigenfrequencies or molecular structure (Figure 9.7).

ATR band distortion and peak shift. The intrinsic absorption (k) (gray, transmission-like) is symmetric about the oscillator position. The ATR band is asymmetric and its maximum is red-shifted relative to the absorption maximum; the distortion grows as the angle approaches the critical angle (39.5° vs 45° on ZnSe). The shift is an optical effect of dispersion and the Fresnel boundary conditions, not a change in the vibrational frequency.

Fig. 9.7
ATR band distortion and peak shift. The intrinsic absorption \(k\) (gray, transmission-like) is symmetric about the oscillator position. The ATR band is asymmetric and its maximum is red-shifted relative to the absorption maximum; the distortion grows as the angle approaches the critical angle (39.5° vs 45° on ZnSe). The shift is an optical effect of dispersion and the Fresnel boundary conditions, not a change in the vibrational frequency.

Beyond simple shifts of the band maximum, ATR spectra frequently exhibit asymmetric line shapes, shoulders, or apparent peak splitting. Such distortions are particularly common for:

  • strong and overlapping absorption bands,

  • polarization-resolved measurements (especially p-polarization),

  • measurements performed close to the critical angle.

Overlapping absorption bands interact through dispersion: the refractive-index change induced by one oscillator modifies the coupling conditions for neighboring oscillators. The resulting intensity redistribution may also occur in normal-incidence transmission spectroscopy. (72)
Overall, ATR spectra may display features that resemble additional chemical bands or band splitting, even when no such modes exist.
A particularly severe form of spectral distortion occurs when the refractive index of the sample approaches—or locally exceeds—the effective critical refractive index for the chosen ATR geometry.
In this regime:
  • the total-reflection condition can be locally violated in spectral regions with strong dispersion,

  • reflectance decreases even where absorption is weak or absent,

  • the logarithmic reflectance (“ATR absorbance”) can increase for purely optical reasons.

Under such conditions, ATR spectra may exaggerate absorption features or suggest absorption where none exists. This behavior is especially relevant for carbonaceous materials, certain metal oxides, and other high-index systems.

Crucially, this situation cannot be diagnosed from the ATR spectrum alone. The measured loss of reflectance does not reveal whether energy is dissipated by absorption or transmitted into the sample due to refractive-index mismatch.

Peak shifts and band-shape distortions have consequences that extend beyond fundamental interpretation. Most infrared spectral databases are based on transmission measurements. Direct comparison of ATR spectra to such databases—without appropriate qualification or correction—can therefore lead to:

  • systematic mismatches in peak positions,

  • reduced identification accuracy,

  • false positives or false negatives in automated matching routines.

These limitations have motivated the development of ATR-specific reference libraries and ATR correction schemes, (69),   (73) as well as practical guidance for recognizing the spectral distortions that distinguish good- from poor-quality ATR measurements, particularly for strong absorption bands compared with transmission. (74) However, without a clear physical understanding, such approaches risk masking distortions rather than resolving them, or worse, introducing apparent agreement for the wrong reasons.
Several recurrent interpretational errors stem from misunderstanding the optical origin of ATR band shifts and distortions:
  • Assigning ATR peak shifts to changes in molecular structure or bonding.

  • Interpreting asymmetric bands or shoulders as evidence of additional chemical species.

  • Treating dispersion-induced distortions as removable “artifacts” without considering geometry and polarization.

Best practice requires that such interpretations be supported by independent evidence rather than inferred from ATR spectra alone.

9.5 Surface and Interfacial Layers: Sensitivity Beyond Penetration Depth

Wave-optical considerations make clear that ATR spectroscopy can be extraordinarily sensitive to nanometer-scale surface and interfacial layers, particularly for p-polarized light. (54) Such layers may arise from chemical modification, contamination, oxidation, mechanical contact, surface roughness, densification, or effective-medium transition regions at the internal reflection element (IRE)–sample interface.

Importantly, the influence of these layers on ATR spectra does not scale with penetration depth and cannot be inferred from bulk absorption arguments. Even when their physical thickness is orders of magnitude smaller than any conventional interaction depth, they can induce substantial spectral changes through boundary-condition–mediated phase–intensity conversion. As a consequence, variations in ATR spectra can be dominated by surface or contact effects rather than by changes in bulk composition, especially in polarization-resolved or angle-dependent measurements.

This sensitivity is not an experimental imperfection. It is an intrinsic consequence of ATR geometry.

The common intuition that ATR probes a “thin layer” of material defined by a penetration depth is misleading, in particular in the presence of interfacial structure. Penetration depth describes a limiting-case field decay in homogeneous, non-absorbing media. It does not determine the sensitivity of reflectance to modifications of boundary conditions.

In reflection geometry, spectral changes arise not from integrating absorption over a path length, but from how the electromagnetic field satisfies boundary conditions at and near the interface. Thin interfacial layers modify these conditions directly. Their influence depends on refractive-index contrast, polarization, and phase relationships—not on the volume of material involved.

As a result, even ultrathin layers can outweigh bulk absorption effects in determining ATR spectral appearance.

Surface and interfacial sensitivity in ATR is strongly polarization dependent. For p-polarized light, the electric-field component normal to the interface couples directly to changes in optical properties across the boundary. This makes p-polarized ATR particularly sensitive to nanometer-scale layers and transition regions.

Angle of incidence further modulates this sensitivity. Near-critical geometries enhance phase sensitivity and the response of the field to changes in the interfacial boundary conditions, amplifying the spectral impact of interfacial structure. Small variations in contact quality, surface composition, or interfacial roughness can therefore produce pronounced and sometimes unexpected spectral changes.

These effects are expected outcomes of reflection optics and should not be dismissed as artifacts.

In practice, interfacial sensitivity may manifest as:

  • apparent band shifts or asymmetry unrelated to bulk chemistry,

  • apparent band splitting, including a blue-shifted component of increased intensity for p-polarization,

  • overall polarization-dependent spectral changes,

  • discrepancies between nominally identical samples measured under slightly different contact pressures or surface states.

Such behavior does not necessarily indicate poor experimental practice. It often reflects genuine sensitivity to interfacial structure.

Because interfacial effects do not scale with penetration depth, they cannot be corrected by simple rescaling, effective-thickness concepts, or bulk-absorption-based ATR correction schemes. Treating surface-dominated variations as bulk chemical changes risks systematic misinterpretation.

From a best-practice perspective:

  • interfacial sensitivity should be anticipated rather than eliminated,

  • polarization and angle dependence should be used diagnostically,

  • correction strategies must respect the boundary-condition origin of the effect, for example through multilayer Fresnel modeling and, where appropriate, dispersion analysis.

Where surface or contact effects are likely to dominate, ATR spectra should be interpreted explicitly as interface-sensitive measurements, not as proxies for bulk absorbance.

9.6 ATR Correction: What Can Be Corrected, What Cannot, and Why

ATR correction is often described as a post-processing step that “converts ATR spectra into transmission-like absorbance spectra.” This description is misleading. In physical terms, ATR correction attempts to invert a wave-optical boundary-value problem under a set of assumptions about the measurement geometry and the optical properties of both the sample and the internal reflection element (IRE).

Whether such an inversion is meaningful—or even possible—depends critically on how the spectrum was measured.

An ATR spectrum records the frequency-dependent reflectance at the IRE–sample interface. This reflectance is governed by Fresnel’s equations and depends on:

  • the complex refractive index of the sample,

  • the refractive index of the IRE,

  • the angle of incidence,

  • the polarization state of the incident radiation,

  • the optical model, e.g. IRE/sample or IRE/layer/sample.

Conceptually, ATR correction therefore consists of two distinct steps:

  • Reconstruction of the sample’s optical response (e.g. complex refractive index or dielectric function) from the measured reflectance.

  • Forward calculation of a desired reference quantity (such as transmission absorbance) from that reconstructed optical response.

Only the second step is straightforward. The first step is fundamentally limited by the information content of the measurement, which is determined entirely by geometry and polarization.

Fresnel’s equations have simple closed forms only for pure s– and p-polarized light. For these polarization states, the reflectance phase can, in principle, be recovered from the reflectance amplitude using causality relations, enabling reconstruction of the complex optical response.

Historically, this is why early ATR experiments were commonly performed with polarizers.

In routine practice today, ATR spectra are often recorded with unpolarized or mixed polarization. This does not remove polarization effects; it merely averages them. The measured spectrum then represents a weighted superposition of s– and p-polarized responses, and phase information required for rigorous inversion is no longer directly accessible.

From a best-practice perspective:

The central consequence is that the apparent absorption of a film can vary with thickness even when its intrinsic optical constants are fixed: band intensities rise and fall as the changing optical thickness shifts the standing wave relative to the absorbing material (Figure 7.4). The same mechanism displaces apparent band positions slightly, because the transmission minima reflect a combination of absorption and interference rather than the intrinsic resonances alone.

  • Ignoring polarization is a choice, not a neutral simplification.

  • That choice limits what can be reconstructed from the data.

Recent work has shown that under specific geometries—most notably near 45° incidence—additional constraints may allow partial reconstruction even without explicit polarizers. However, this remains conditional on knowing the polarization state.

The angle of incidence (AOI) is one of the most influential parameters in ATR spectroscopy, yet it is rarely reported explicitly.

The AOI determines:

The central consequence is that the apparent absorption of a film can vary with thickness even when its intrinsic optical constants are fixed: band intensities rise and fall as the changing optical thickness shifts the standing wave relative to the absorbing material (Figure 7.4). The same mechanism displaces apparent band positions slightly, because the transmission minima reflect a combination of absorption and interference rather than the intrinsic resonances alone.

  • the strength of the signal,

  • sensitivity to refractive-index changes,

  • and whether the total-reflection condition is satisfied across the spectrum.

Because the refractive index of the sample is frequency dependent, the critical angle is also frequency dependent. Even if total reflection is satisfied in transparent regions, it may be locally violated near strong absorption bands due to dispersion.

When this occurs:

  • correction becomes ill-conditioned when an appreciable spectral region is affected,

  • ATR absorbance can increase even in the absence of genuine absorption,

  • and correction methods based on weak-absorption assumptions fail.

The IRE material is not a neutral component. Its refractive index sets the boundary conditions of the experiment and strongly influences ATR spectra.

Using a high-index IRE:

  • increases robustness against violations of the total-reflection condition,

  • reduces overall signal intensity,

  • suppresses optics-induced peak shifts and distortions.

Accordingly, using a lower-index IRE:

  • increases sensitivity,

  • but amplifies dispersion-induced distortions and non-linearities.

ATR absorbance spectra recorded with different IRE materials are therefore not equivalent measurements and should not be compared directly without accounting for this difference. Most commercially available ATR accessories use comparatively lower-index IRE materials such as ZnSe or diamond and an angle of incidence near 45°. This produces comparatively strong signals because the geometry is close to the critical-angle regime for many substances.

A key quantity governing ATR behavior is the refractive index of the sample outside absorption bands, often denoted \(n_\infty\). This parameter determines the baseline optical contrast at the interface and strongly influences:

  • the effective critical angle,

  • coupling efficiency into the sample,

  • the magnitude of dispersion-induced distortions.

Importantly, \(n_\infty\) is not reliably retrievable from ATR spectra alone. It must therefore be regarded as an external boundary condition, not as a fitting parameter to be absorbed implicitly into correction schemes. In principle, sufficiently well-calibrated polarized reflectance data or subcritical reflection measurements can constrain the background refractive index; but in routine ATR correction \(n_\infty\) should be supplied from independent information rather than treated as an unconstrained fit parameter.

Neglecting or misestimating \(n_\infty\) is a frequent source of apparent inconsistencies concerning oscillator position and strength in ATR interpretation and correction.

In principle, physically grounded ATR correction can address:

  • dispersion-induced peak shifts,

  • geometry-dependent intensity scaling,

  • peak shape changes.

These corrections rely on explicit assumptions about polarization, AOI, IRE properties, and, depending on the method, the weak-absorption regime.

No correction method—regardless of sophistication—can compensate for missing information. In particular, ATR correction cannot uniquely resolve:

  • violations of total reflection without polarization knowledge,

  • separation of absorption losses from transmission losses below the critical angle,

  • effects of strong scattering or poor optical contact, and

  • in general, deviations from the assumed optical/geometric model.

Correction results obtained under such conditions must therefore be interpreted with caution, even if they appear numerically plausible.

A subtler point is that not every non-linearity in an ATR spectrum is an artifact to be corrected away. It is useful to separate two kinds. 

Geometry-induced non-linearities arise from the IRE refractive index, the incidence angle, the polarization, the interface structure, and the reflectance-to-absorbance transformation; they are extrinsic, in the sense that measuring the same sample with a different element or angle changes them, and removing them is the legitimate aim of correction. 

Material non-linearities — mixture effects, local-field effects, density changes, molecular interactions, and heterogeneity — are different: they make the optical constants themselves non-linear functions of composition or state, and they belong to the sample, not to the geometry.

A correction method should not erase them merely because a linear calibration model would prefer them gone. The sound division of labor is for optical modeling to build the cleanest possible material feature space — a complex refractive index, a dielectric function, or a corrected absorption-related quantity that separates geometry from material response — and for chemometrics to operate on that feature space afterward.

This also supplies the test of whether a correction has succeeded. A corrected \(n,k\) spectrum is convincing not because it looks smooth or transmission-like, but because, when inserted back into the forward Fresnel model, it reproduces the measured reflectance with small residuals. A correction that cannot pass this forward-model check has not solved the underlying optical problem, however plausible its output appears. Residuals and diagnostics are therefore part of the meaning of ATR correction, not optional extras.

ATR correction is not a universal remedy that turns ATR spectra into “true absorbance.”

It is a conditional reconstruction whose validity depends on:

  • measurement geometry,

  • polarization control,

  • refractive-index contrast,

  • and the absorption regime.

Used with physical awareness, ATR correction can be powerful. Used indiscriminately, it risks obscuring rather than revealing the underlying optical response.

The following table compares the main families of ATR correction method — from simple wavelength scaling and Kramers–Kronig-based absorbance correction, through KKR- and Fresnel-based determination of the optical constants, to full dispersion analysis — by their underlying principle, limitations and restrictions, speed, additional requirements, and key references. It is intended as an orientation aid rather than a prescriptive checklist, and its entries must be interpreted in the context of the specific optical geometry and material system discussed above.

Table 9.1.
Correction methods for ATR spectra.
 Methods are ordered roughly from the simplest scaling to full dispersion analysis. The supercritical condition is written with the chapter’s convention (\(n_1\) = IRE, \(n_2\) = sample); “safely supercritical” means \(n_{\text{sample}} < n_{\text{IRE}}\sin\alpha\) holds over the relevant spectral range.

Method Principle Limitations and restrictions Speed Additional requirements References
Wavelength correction Multiply ATR absorbance by the wavenumber Dispersion and absorption neglected Very fast (< 0.1 s) — 75
PMATRC (Poor Man’s ATR correction) Wavelength scaling plus a Kramers–Kronig-transformed absorbance correction Relative peak intensities are not corrected Very fast (< 0.1 s) Polarization angle 76
Effective-depth correction Optical-constant function via KKR (iterative), starting from an estimated effective depth Safely supercritical geometry (nsample < nIRE sin α) over the whole spectral range; n∞ must be known; initial estimate uses the low-absorption assumption Fast (< 1 s) Polarization angle 77, 78, 79
Absorbance correction (after Bertie & Eysel) Optical-constant function via KKR (iterative), starting from an estimated absorption index Safely supercritical geometry (nsample < nIRE sin α) over the whole spectral range; n∞ must be known; initial estimate uses the low-absorption assumption Fast (< 1 s) Polarization angle 79, 80, 81
Fresnel-equation iterative correction Iterative determination of the optical constants without the low-absorption assumption The Fresnel forward model can describe both subcritical and supercritical regimes, but the inversion remains conditional on geometry, polarization, and information content; n∞ must be known; best for medium–strong oscillators Fast (< 1 s) Polarization angle 70
Sophisticated Fresnel/KKR inversion (direct) Optical-constant function via KKR (direct, non-iterative) Safely supercritical geometry (nsample < nIRE sin α) at the band edges of the spectral range; n∞ must be known Fast (< 1 s) Polarization angle 69, 82
Dispersion analysis Optical-constant function via causal dispersion relations (oscillator model, iterative) n∞ must be known; requires a physically plausible, causal oscillator model with parameter plausibility and residual diagnostics Very slow (10–100 min) Polarization angle; experienced operator 83, 84

9.7 Best-Practice Criteria for ATR Spectroscopy

The following points are not rigid rules. They define the conditions under which ATR spectra can be interpreted, corrected, and compared in a physically meaningful way.

ATR absorbance spectroscopy is particularly well suited when:

  • sample preparation for transmission is impractical or undesirable,

  • the sample is optically dense but has moderate refractive index,

  • surface sensitivity is desired while bulk optical properties remain relevant,

  • for robust bulk quantification, the measurement geometry avoids uncontrolled proximity to critical-angle conditions across the spectral range.

Under these conditions, ATR provides robust, reproducible spectra with high practical value.

Direct interpretation of ATR absorbance spectra (without correction) is defensible when:

  • absorption bands are weak to moderate,

  • the refractive index of the sample remains well below that of the IRE,

  • peak positions and intensities are not over-interpreted as molecular constants,

  • comparisons are restricted to spectra recorded under identical ATR conditions, and any concentration or composition series has been independently checked for linearity.

In such cases, ATR spectra should be described explicitly as ATR reflectance-derived spectra (“ATR absorbance”), not as transmission absorbance equivalents.

ATR correction becomes necessary when:

  • quantitative comparison with true absorbance spectra is intended,

  • concentration determination is required beyond empirical calibration,

  • strong bands exhibit pronounced shifts or distortions,

  • spectra recorded with different IREs, angles of incidence, or polarizations are compared.

For reproducibility and critical evaluation, authors should report at least:

  • the IRE material,

  • the accessory type or number of reflections,

  • the nominal angle of incidence (and angular spread, if known),

  • the polarization state or the effective-polarization assumption,

  • the reference/background procedure,

  • the sample contact protocol (e.g. contact pressure),

  • the spectral range,

  • the independent \(n_\infty\) value and its source, if correction or inversion is used,

  • whether and which type of ATR correction was performed,

  • known limitations of the chosen approach.

Failure to report these parameters limits physical interpretation and reuse of the data.

Best practice in quantitative ATR spectroscopy recognizes that absorption alone is often insufficient. Where possible, analyses should combine:

  • refractive-index information, in particular the refractive index at the higher wavenumber end of the spectrum \(n_\infty\). Accurate values of \(n_\infty\) are required for sophisticated ATR correction.

  • geometry-specific parameters (IRE, angle of incidence, polarization).

This multi-observable approach improves robustness, reduces non-linearity, and enhances transferability across instruments and configurations.

ATR spectra are modality-specific representations, not universal chemical fingerprints.

As a result:

  • chemometric and AI models trained on ATR absorbance data implicitly learn geometry and IRE dependencies,

  • model transfer across instruments or ATR accessories without retraining is unsafe,

  • blind preprocessing may remove physically meaningful structure.

Responsible use of ATR data in automated workflows requires explicit documentation of ATR geometry and, where possible, inclusion of refractive-index-related descriptors. For physics-informed complex-valued chemometrics, where feasible, the use of physically meaningful feature spaces — such as the complex refractive index \(n, k\), the dielectric function \(\varepsilon\), or a corrected absorption-related quantity — is recommended instead of ATR absorbance spectra.

ATR spectroscopy is neither a poor substitute for transmission nor a purely surface-selective curiosity. It is a physically rich measurement modality whose spectra encode absorption, dispersion, and boundary conditions simultaneously.

Responsible use of ATR spectroscopy requires:

  • awareness of its wave-optical foundations,

  • explicit reporting of measurement parameters,

  • recognition of the limits of correction,

  • and careful distinction between intrinsic optical effects and experimental artifacts.

When these conditions are met, ATR spectroscopy becomes a powerful and quantitative tool. When they are ignored, ATR spectra may appear deceptively simple while hiding systematic distortions.

References

  1. Tolstoy VP, Chernyshova I, Skryshevsky VA. Handbook of Infrared Spectroscopy of Ultrathin Films. Wiley; 2003. doi:10.1002/047123432X
  1. Harrick NJ. Internal Reflection Spectroscopy. Interscience Publishers; 1967.
  1. Hansen WN. Expanded formulas for attenuated total reflection and the derivation of absorption rules for single and multiple ATR spectrometer cells. Spectrochim Acta. 1965;21(4):815-833. doi:10.1016/0371-1951(65)80039-X

65. Mayerhöfer TG, Popp J. Understanding the Role of the Evanescent Field in Attenuated Total Reflection (ATR) Spectroscopy. Appl Spectrosc. 2026;80(2):125-132. doi:10.1177/00037028251358400

  1. Fahrenfort J. Attenuated total reflection. Spectrochim Acta. 1961;17(&):698-709. doi:10.1016/0371-1951(61)80136-7
  1. Milosevic M. Internal Reflection and ATR Spectroscopy. John Wiley & Sons; 2012. doi:10.1002/9781118309742
  1. Mayerhöfer TG, Popp J. Electric field standing wave effects in internal reflection and ATR spectroscopy. Spectrochim Acta A Mol Biomol Spectrosc. 2018;191:165-171. doi:10.1016/j.saa.2017.10.007
  1. Mayerhöfer TG, Popp J. Sophisticated ATR correction at arbitrary incidence and polarization. in preparation. 2026;
  1. Mayerhöfer TG, Popp J. Attenuated total and internal reflection correction based on Fresnel’s equations: beyond the low absorption assumption. Appl Spectrosc. 2025;79(12):1747-1757. doi:10.1177/00037028251352565
  1. Mayerhöfer TG, Ilchenko O, Kutsyk A, Popp J. Beyond Beer’s Law: quasi-ideal binary liquid mixtures. Appl Spectrosc. 2022;76(1):92-104. doi:10.1177/00037028211056293
  1. Mayerhöfer TG, Pahlow S, Ivanovski V, Popp J. Dispersion related coupling effects in IR spectra on the example of water and amide I bands. Spectrochim Acta A Mol Biomol Spectrosc. 2023;288:122115. doi:10.1016/j.saa.2022.122115
  1. Mirabella FM. Internal reflection spectroscopy. Appl Spectrosc Rev. 1985;21(1-2):45-178. doi:10.1080/05704928508060428
  1. Miseo EV, Larkin PJ. Fourier transform infrared spectroscopy (FT-IR) diamond attenuated total reflection (ATR) measurements: The good, the bad, and the (really) ugly. Appl Spectrosc Pract. 2025;3(2). doi:10.1177/27551857251336262
  1. Mayerhöfer TG, Popp J. Understanding advanced attenuated total reflection correction: The low absorbance assumption. Appl Spectrosc. 2025;79(2):298-305. doi:10.1177/00037028241268024
  1. Mayerhöfer TG, Costa WDP, Popp J. A simplified, effective ATR correction method for spectral identification. Spectrochim Acta A Mol Biomol Spectrosc. 2025;341:126425. doi:10.1016/j.saa.2025.126425
  1. Mayerhöfer TG, Popp J. Developing correction methods by revisiting the concept of effective thickness in attenuated total reflection spectroscopy. Appl Spectrosc. 2025;79(3):465-472. doi:10.1177/00037028241290838
  1. Ohta K, Ishida H. Efficient method for optical constant determination by FTIR-ATR. In: Heise HM, Korte EH, Siesler HW, eds. 8th International Conference on Fourier Transform Spectroscopy. Vol 1575. SPIE; 1992:604-605. doi:10.1117/12.56372
  1. Mayerhöfer TG, Popp J. Understanding and employing (non-)linearities in attenuated total reflection spectroscopy. Appl Spectrosc. 2025;79(8):1290-1296. doi:10.1177/00037028251317540
  1. Bertie JE, Eysel HH. Infrared intensities of liquids I: Determination of infrared optical and dielectric constants by FT-IR using the CIRCLE ATR cell. Appl Spectrosc. 1985;39(3):392-401. doi:10.1366/0003702854248575
  1. Bertie JE, Zhang SL, Manji R. Infrared intensities of liquids X: Accuracy of current methods of obtaining optical constants from multiple attenuated total reflection measurements using the CIRCLE cell. Appl Spectrosc. 1992;46(11):1660-1665. doi:10.1366/0003702924926844
  1. Mayerhöfer TG, Costa WDP, Popp J. Sophisticated attenuated total reflection correction within seconds for unpolarized incident light at 45°. Appl Spectrosc. 2024;78(3):321-328. doi:10.1177/00037028231219528

83. Milosevic M, Wendland N, Lee RE, Gregory BW. The usefulness of spectroscopic simulations. Appl Spectrosc. 2020;74(3):305-313. doi:10.1177/0003702819893689

  1. Vieira LG. Infrared optical functions of water retrieved using attenuated total reflection spectroscopy. Appl Spectrosc. 2023;77(2):178-186. doi:10.1177/00037028221128813

The Field Guide is a living document. Questions, corrections, and suggestions for this chapter are very welcome — join the discussion in the Coblentz Field Guide forum.